ABSTRACT
BACKGROUND:
One of the key issues in electroencephalogram (EEG) monitoring is accurate signal acquisition with less cumbersome electrodes. In this study, the L2 phase electro-deposited nanoporous platinum (L2-ePt) electrode is introduced, which is a new type of electrode that utilizes a stable nanoporous platinum surface to reduce the skin-electrode impedance.METHODS:
L2-ePt electrodes were fabricated using electro-deposition technique. Then, the effect of the nanoporous surface on the surface roughness and the electrode impedance were observed from the L2-ePt electrodes and the flat platinum (FlatPt) electrode. The skin-electrode impedances of the L2-ePt electrodes, a gold cup electrode, and the FlatPt electrode were evaluated when placed on the hairy occipital area of the head in ten subjects. For the validation of using the L2-ePt electrode, a correlational analysis of the alpha rhythms was performed in the same subjects for simultaneous EEG recordings using the L2-ePt and clinically-used EEG electrodes.RESULTS:
The results indicated that the L2-ePt electrode with a roughness factor of 200 had the lowest mean impedance performance. Moreover, the proposed L2-ePt electrode showed a significantly lower mean skin-electrode impedance than the FlatPt electrode. Finally, the EEG signal quality recorded by the L2-ePt electrode (r = 0.94) was comparable to that of the clinically-used gold cup electrode.CONCLUSION:
Based on these results, the proposed L2-ePt electrode is suitable for use in various high-quality EEG applications.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Assunto principal:
Platina
/
Impedância Elétrica
/
Eletrodos
/
Eletroencefalografia
/
Ritmo alfa
/
Cabeça
Idioma:
Inglês
Revista:
Journal of Korean Medical Science
Ano de publicação:
2018
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS