Comparison of Ocular Biometry Using New Swept-source Optical Coherence Tomography-based Optical Biometer with Other Devices
Korean Journal of Ophthalmology ; : 257-264, 2018.
Artigo em Inglês | WPRIM | ID: wpr-716264
ABSTRACT

PURPOSE:

To evaluate the agreement between optical biometry with swept-source optical coherence tomography-based optical biometry (IOLMaster 700) and other devices.

METHODS:

A total of 137 eyes (78 patients) with cataracts were included in this retrospective study. Axial length (AL), anterior chamber depth (ACD), keratometry, and white-to-white (WTW) distance measured using IOLMaster 700 were compared with results for the following five different biometers IOLMaster 500, A-scan, automated refractor, manual keratometry, and Galilei G4. Differences and correlations among the devices were assessed using the Bland-Altman plot and intraclass correlation coefficient (ICC).

RESULTS:

For AL values, the IOLMaster 700, IOLMaster 500, and A-scan measurements showed excellent agreement (all ICC >0.99). For ACD values, ICC of IOLMaster 700 and Galilei G4 was 0.965 but A-scan was poorly correlated with either IOLMaster 700 or Galilei G4. The ICCs of IOLMaster 700 and other devices were all greater than 0.9 for average keratometry, but those of the mean cylinder keratometry were all between 0.7 and 0.8. The mean difference in the WTW distance between the IOLMaster 700 and Galilei G4 was 0.029 mm, but the ICC was 0.525. AL measurements were not possible for 10 eyes with the IOLMaster 500 but were obtained in all eyes with the IOLMaster 700.

CONCLUSIONS:

In clinical practice, AL, ACD, and average keratometry values of IOLMaster 700 can be used interchangeably with those of the other devices tested. However, the ACD value between IOLMaster 700 and A-scan or the WTW distance between IOLMaster 700 and Galilei G4 are not interchangeable because of clinical and statistical differences in measurements between the devices.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Catarata / Estudos Retrospectivos / Biometria / Cimetidina / Câmara Anterior Tipo de estudo: Guia de Prática Clínica / Estudo observacional Idioma: Inglês Revista: Korean Journal of Ophthalmology Ano de publicação: 2018 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Catarata / Estudos Retrospectivos / Biometria / Cimetidina / Câmara Anterior Tipo de estudo: Guia de Prática Clínica / Estudo observacional Idioma: Inglês Revista: Korean Journal of Ophthalmology Ano de publicação: 2018 Tipo de documento: Artigo