Your browser doesn't support javascript.
loading
Clinical Usefulness of ImmuneCheck™ IgG for Rapid Semiquantitation of Total IgG
Laboratory Medicine Online ; : 140-147, 2018.
Artigo em Coreano | WPRIM | ID: wpr-717397
ABSTRACT

BACKGROUND:

Conventional IgG assays require costly equipment and skilled experts. Semiquantitative measurement of total IgG using point-of-care testing devices may be the solution for these limitations. This study evaluated the reproducibility of the ImmuneCheck™ IgG assay (ProteomeTech Inc., Korea) and the correlation of its results with conventional laboratory IgG results in the serum and whole blood.

METHODS:

Both the serum and whole blood samples from 120 patients were used. To evaluate the intra-test reproducibility and inter-test correlation, intraclass correlation coefficient (ICC) analysis was used.

RESULTS:

The concentration of serum total IgG measured by cobas® 6000 (Roche Diagnostics, Switzland) ranged from 690.4 to 2,756.4 mg/dL. The intra-test reproducibility of ImmuneCheck™ IgG was high (Serum ICC=0.724, P < 0.001; Whole blood ICC=0.843, P < 0.001). The inter-test correlation between the ImmuneCheck™ IgG and cobas® 6000 results was very good (Serum ICC=0.805, P < 0.001; Whole blood ICC=0.842, P < 0.001). Because there were no samples with a total IgG level lower than 600 mg/dL, the pre-existing serum samples were diluted and then the linearity tests were conducted. The intra-test reproducibility for the diluted serum samples was almost perfect (ICC=0.995, P < 0.001), and the inter-test correlation between the ImmuneCheck™ IgG and cobas® 6000 results was also strong (ICC=0.992, P < 0.001).

CONCLUSIONS:

The ImmuneCheck™ IgG assay is reproducible and highly correlated with the conventional IgG assay for the serum and whole blood. It could be applied for the rapid detection of total IgG.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Imunoglobulina G / Testes Imediatos Tipo de estudo: Estudo diagnóstico Limite: Humanos Idioma: Coreano Revista: Laboratory Medicine Online Ano de publicação: 2018 Tipo de documento: Artigo