Estimation of RTP Accuracy Based the International Reference Level / 의학물리
Korean Journal of Medical Physics
;
: 113-119, 2008.
Artigo
em Coreano
| WPRIM
| ID: wpr-7199
ABSTRACT
In this study, we have estimated error of calculation results for 6-type RTP systems and investigated a toleration for error of the RTPs referenced from the evaluation items of AAPM Report-62. For this study, we have introduced the concept of 'normal dose rate(NDR)' and compared the results of experiment and calculation from RTPs at the same reference level. The results from all RTPs were satisfied at various field shapes and heterogeneous phantom materials except the surface irregularity.
Texto completo:
DisponíveL
Índice:
WPRIM (Pacífico Ocidental)
Idioma:
Coreano
Revista:
Korean Journal of Medical Physics
Ano de publicação:
2008
Tipo de documento:
Artigo
Similares
MEDLINE
...
LILACS
LIS