Development of A Surface Acoustic Wave LB MembraneImmunity Sensor / 华中科技大学学报(医学)(英德文版)
Article em Zh | WPRIM | ID: wpr-735641
Biblioteca responsável: WPRO
ABSTRACT
A new kind of biosensor for immunology was developed by ultrasonic technique and LB membrane. A double delay-line resonator was made by using ST-cut quartz crystal with working frequency of 149.7 MHz. Then a layer of LB membrane was covered on it. When anti-IgM antibody of various concentrations was added to it, the sensor can be used to detect IgM antigen. The biosensor was highly sensitive, small and light. The experimental results showed that the working frequency change of the sensor was proportional to the concentration of antibody with its dilution ratio between 1:10000 and 1:100. It was also first observed that the frequency curve of the sensor resulting from the reaction of IgM antigen and antibody undulated in the experiment.
Palavras-chave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: Journal of Huazhong University of Science and Technology (Medical Sciences) Ano de publicação: 2000 Tipo de documento: Article