Optimization of exposure parameters and relationship between subjective and technical image quality in cone-beam computed tomography
Imaging Science in Dentistry ; : 139-151, 2019.
Artigo em Inglês | WPRIM | ID: wpr-763991
ABSTRACT

PURPOSE:

This study was performed to investigate the effect of exposure parameters on image quality obtained using a cone-beam computed tomography (CBCT) scanner and the relationship between physical factors and clinical image quality depending on the diagnostic task. MATERIALS AND

METHODS:

CBCT images of a SedentexCT IQ phantom and a real skull phantom were obtained under different combinations of tube voltage and tube current (Alphard 3030 CBCT scanner, 78–90 kVp and 2–8 mA). The images obtained using a SedentexCT IQ phantom were analyzed technically, and the physical factors of image noise, contrast resolution, spatial resolution, and metal artifacts were measured. The images obtained using a real skull phantom were evaluated for each diagnostic task by 6 oral and maxillofacial radiologists, and each setting was classified as acceptable or unacceptable based on those evaluations. A statistical analysis of the relationships of exposure parameters and physical factors with observer scores was conducted.

RESULTS:

For periapical diagnosis and implant planning, the tube current of the acceptable images was significantly higher than that of the unacceptable images. Image noise, the contrast-to-noise ratio (CNR), the line pair chart on the Z axis, and modulation transfer function (MTF) values showed statistically significant differences between the acceptable and unacceptable image groups. The cut-off values obtained using receiver operating characteristic curves for CNR and MTF 10 were useful for determining acceptability.

CONCLUSION:

Tube current had a major influence on clinical image quality. CNR and MTF 10 were useful physical factors that showed significantly associations with clinical image quality.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Crânio / Curva ROC / Artefatos / Diagnóstico / Tomografia Computadorizada de Feixe Cônico / Ruído Tipo de estudo: Estudo diagnóstico / Estudo prognóstico Idioma: Inglês Revista: Imaging Science in Dentistry Ano de publicação: 2019 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Crânio / Curva ROC / Artefatos / Diagnóstico / Tomografia Computadorizada de Feixe Cônico / Ruído Tipo de estudo: Estudo diagnóstico / Estudo prognóstico Idioma: Inglês Revista: Imaging Science in Dentistry Ano de publicação: 2019 Tipo de documento: Artigo