Ultrashort Echo Time MRI (UTE-MRI) Quantifications of Cortical Bone Varied Significantly at Body Temperature Compared with Room Temperature
Investigative Magnetic Resonance Imaging ; : 202-209, 2019.
Artigo em Inglês | WPRIM | ID: wpr-764185
ABSTRACT

PURPOSE:

To investigate the temperature-based differences of cortical bone ultrashort echo time MRI (UTE-MRI) biomarkers between body and room temperatures. Investigations of ex vivo UTE-MRI techniques were performed mostly at room temperature however, it is noted that the MRI properties of cortical bone may differ in vivo due to the higher temperature which exists as a condition in the live body. MATERIALS AND

METHODS:

Cortical bone specimens from fourteen donors (63 ± 21 years old, 6 females and 8 males) were scanned on a 3T clinical scanner at body and room temperatures to perform T1, T2*, inversion recovery UTE (IR-UTE) T2* measurements, and two-pool magnetization transfer (MT) modeling.

RESULTS:

Single-component T2*, IR-T2*, short and long component T2*s from bi-component analysis, and T1 showed significantly higher values while the noted macromolecular fraction (MMF) from MT modeling showed significantly lower values at body temperature, as compared with room temperature. However, it is noted that the short component fraction (Frac1) showed higher values at body temperature.

CONCLUSION:

This study highlights the need for careful consideration of the temperature effects on MRI measurements, before extending a conclusion from ex vivo studies on cortical bone specimens to clinical in vivo studies. It is noted that the increased relaxation times at higher temperature was most likely due to an increased molecular motion. The T1 increase for the studied human bone specimens was noted as being significantly higher than the previously reported values for bovine cortical bone. The prevailing discipline notes that the increased relaxation times of the bound water likely resulted in a lower signal loss during data acquisition, which led to the incidence of a higher Frac1 at body temperature.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Relaxamento / Doadores de Tecidos / Temperatura Corporal / Imageamento por Ressonância Magnética / Água / Biomarcadores / Incidência Tipo de estudo: Estudo de incidência / Estudo prognóstico Limite: Feminino / Humanos Idioma: Inglês Revista: Investigative Magnetic Resonance Imaging Ano de publicação: 2019 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Relaxamento / Doadores de Tecidos / Temperatura Corporal / Imageamento por Ressonância Magnética / Água / Biomarcadores / Incidência Tipo de estudo: Estudo de incidência / Estudo prognóstico Limite: Feminino / Humanos Idioma: Inglês Revista: Investigative Magnetic Resonance Imaging Ano de publicação: 2019 Tipo de documento: Artigo