Performance Evaluation of Infusion Systems for Red Blood Cell Transfusion
Journal of Laboratory Medicine and Quality Assurance ; : 161-165, 2019.
Artigo em Coreano | WPRIM | ID: wpr-765647
ABSTRACT

BACKGROUND:

Transfusion in neonates and infants can be performed using an electromechanical infusion system that has appropriate accuracy in terms of flow rate, volume, and bolus. However, there are no infusion systems approved for transfusion in Korea. In this study, we evaluate the performance of two electromechanical infusion systems for transfusion in pediatric patients.

METHODS:

We tested two systems, Baxter and Terumo, using 9 units of leukocyte-filtered red blood cells. The blood samples were delivered through the systems at constant speeds of 10, 30, and 100 mL/hr, and the accuracy in terms of the delivered volume was estimated. Before and after infusion, hemoglobin, hematocrit, plasma hemoglobin, potassium, and lactate dehydrogenase levels were measured in each sample. The percentage of hemolysis (%Hemolysis) was calculated to evaluate the safety of the infusion systems.

RESULTS:

For Terumo, the mean error rate of the infused volume was less than 5%. We expect that Terumo can transfuse blood at a volume close to the set volume. Further, both infusion systems showed acceptable %Hemolysis levels (mean±standard deviation Terumo, 0.14±0.04; Baxter, 0.17±0.06).

CONCLUSIONS:

Both infusion systems can be used safely for transfusion in pediatric patients.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Plasma / Potássio / Transfusão de Eritrócitos / Eritrócitos / Hematócrito / Hemólise / Coreia (Geográfico) / L-Lactato Desidrogenase Limite: Humanos / Lactente / Recém-Nascido País/Região como assunto: Ásia Idioma: Coreano Revista: Journal of Laboratory Medicine and Quality Assurance Ano de publicação: 2019 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Plasma / Potássio / Transfusão de Eritrócitos / Eritrócitos / Hematócrito / Hemólise / Coreia (Geográfico) / L-Lactato Desidrogenase Limite: Humanos / Lactente / Recém-Nascido País/Região como assunto: Ásia Idioma: Coreano Revista: Journal of Laboratory Medicine and Quality Assurance Ano de publicação: 2019 Tipo de documento: Artigo