Your browser doesn't support javascript.
loading
Development of a Portable Pulse Oxymetry Simulator System / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 345-347, 2019.
Artigo em Chinês | WPRIM | ID: wpr-772489
ABSTRACT
This article describes the design of a portable blood oxygen simulation system that can be used to simulate various blood gas saturation states of the human body. The system can be used to simulate various states of blood gas saturation, and can also simulate large blood oxygen saturation dynamic range, pulse rate range and perfusion index range. It can be used for testing, but not for clinical examination instruments. Moreover, the system has the characteristics of small size and low cost compared with the commercial blood oxygen simulator. Although the simulation system is not directly used for the detection of blood gas saturation of patients, it is also an essential equipment in the production and testing process, so it has certain practical value.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Oxigênio / Oximetria / Frequência Cardíaca Limite: Humanos Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2019 Tipo de documento: Artigo