Application of self-made fine-tuning device in distal locking process of femoral interlocking intramedullary nail / 中国骨伤
China Journal of Orthopaedics and Traumatology
; (12): 161-165, 2019.
Article
em Zh
| WPRIM
| ID: wpr-776118
Biblioteca responsável:
WPRO
ABSTRACT
OBJECTIVE@#To evaluate the curative efficacy of self-made fine-tuning setting in the process of femoral distal implantation of intramedullary nail.@*METHODS@#From October 2015 to October 2017, 66 cases of femoral shaft fracture were treated with anterograde interlocking intramedullary nail including 45 males and 21 females with a mean age of(37.21±11.18) years old. Among them, 36 cases were treated with the manufacture's aiming device and self-made fine-tuning setting (research group), other 30 cases were treated with the manufacture's aiming device(control group). The mean operation time, the times of C-arm scan in surgery, the post-operation complications and the fracture union were observed and compared in two groups.@*RESULTS@#Sixty-two cases acquired 8 to 15 months with a mean time of 12.4 months follow-up visit. The post-operation complications and the fracture union between the two groups had no significant difference(>0.05), the mean operation time and the times of C-arm scan in surgery had statistically significant difference(<0.05).@*CONCLUSIONS@#Self-made fine-tuning setting in the process of femoral interblocking intramedullary nail could shorten operation time and reduce the the times of C-arm scan.
Palavras-chave
Texto completo:
1
Índice:
WPRIM
Assunto principal:
Período Pós-Operatório
/
Pinos Ortopédicos
/
Resultado do Tratamento
/
Fraturas do Fêmur
/
Duração da Cirurgia
/
Fixação Intramedular de Fraturas
Limite:
Adult
/
Female
/
Humans
/
Male
Idioma:
Zh
Revista:
China Journal of Orthopaedics and Traumatology
Ano de publicação:
2019
Tipo de documento:
Article