Your browser doesn't support javascript.
loading
Magnetic Resonance Imaging Compatibility of the Polymer-based Cochlear Implant
Clinical and Experimental Otorhinolaryngology ; : S19-S23, 2012.
Artigo em Inglês | WPRIM | ID: wpr-77641
ABSTRACT

OBJECTIVES:

In this study, we compared the magnetic resonance (MR) image artifacts caused by a conventional metal-based cochlear implant and a newly developed liquid crystal polymer (LCP)-based device.

METHODS:

The metal-based cochlear implant system (Nurobiosys Co.) was attached to side of the head of a subject and the LCP-based device was attached to opposite side. In both devices, alignment magnets were removed for safety. Magnetic resonance imaging (MRI) was performed on a widely used 3.0 T and an ultra-high 7.0 T MRI machine. 3.0 and 7.0 T MR images were acquired using T1- and T2*-weighted gradient echo sequences, respectively.

RESULTS:

In the 3.0 T images, the metal-based device on the left side generated the significant amount of artifacts. The MR images in the proximity of the metal package were obscured by the artifacts in both axial and sagittal views. On the other hand, the MR images near the LCP-based device were relatively free from the artifacts and clearly showed the brain structures. 7.0 T MR images showed the more severe distortion in the both sides but the metal-based cochlear implant system caused a much larger obscure area than the LCP-based system.

CONCLUSION:

The novel LCP-based cochlear implant provides a good MRI compatibility beyond present-day cochlear implants. Thus, MR images can be obtained from the subjects even with the implanted LCP-based neural prosthetic systems providing useful diagnostic information. Furthermore, it will be also useful for functional MRI studies of the auditory perception mechanism after cochlear implantations as well as for positron emission tomography-MRI hybrid imaging.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Polímeros / Percepção Auditiva / Encéfalo / Imageamento por Ressonância Magnética / Espectroscopia de Ressonância Magnética / Quimera / Implantes Cocleares / Artefatos / Implante Coclear / Elétrons Idioma: Inglês Revista: Clinical and Experimental Otorhinolaryngology Ano de publicação: 2012 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Polímeros / Percepção Auditiva / Encéfalo / Imageamento por Ressonância Magnética / Espectroscopia de Ressonância Magnética / Quimera / Implantes Cocleares / Artefatos / Implante Coclear / Elétrons Idioma: Inglês Revista: Clinical and Experimental Otorhinolaryngology Ano de publicação: 2012 Tipo de documento: Artigo