Your browser doesn't support javascript.
loading
The three dimensional finite element analysis of the stress distribution according to the implant and thread designs
Journal of the Korean Association of Maxillofacial Plastic and Reconstructive Surgeons ; : 443-452, 2004.
Artigo em Coreano | WPRIM | ID: wpr-784577
ABSTRACT
biologic load to disperse the load adequately to the surrounding bone. In this study, based upon the implant and thread designs, we have assessed the equivalent stress and equivalent strain in the cortical and cancellous bone with finite element analysis when under 100N vertical force and 20N horizontal force were simultaneously loaded on the following 6 different kinds of implant design and 4 different kinds of thread design. According to the results, the cylinder type or cylinder and tapered type is the most favorable design in the same diameter. To supplement the shortcoming of the tapered type and stepped cylinder type in the respect of design and project area, it is better to increase the diameter of the implants. The square and reverse threads are better than triangular and plateau thread in the respect of maximum equivalent stress and strain in the cortical and cancellous bone and project area.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Análise de Elementos Finitos Idioma: Coreano Revista: Journal of the Korean Association of Maxillofacial Plastic and Reconstructive Surgeons Ano de publicação: 2004 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Análise de Elementos Finitos Idioma: Coreano Revista: Journal of the Korean Association of Maxillofacial Plastic and Reconstructive Surgeons Ano de publicação: 2004 Tipo de documento: Artigo