Feasibility of finite element analysis in the application of quality inspection on implantable medical devices / 医用生物力学
Journal of Medical Biomechanics ; (6): E001-E006, 2014.
Artigo em Chinês | WPRIM | ID: wpr-804356
ABSTRACT
For implantable medical devices, traditional mechanical property testing is achieved by mechanical testing devices, while such method is lack of pertinence during sampling of series of products, and also time consuming during experimental testing. With the complete development of finite element analysis (FEA) techniques, introducing FEA to quality supervision and inspection will become a scientific method for saving resources and time, and also improve the pertinence. In this study, the creditability of FEA in application of highest risk sample selection, failure analysis of marketed products and auxiliary optimization of fatigue test testing was verified through case study and experimental testing. The feasibility of FEA during implants inspection was illustrated. In order to ensure the accuracy and reliability of FEA in the application of implants inspection, the essentials of establishing related finite element standards were proposed.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Journal of Medical Biomechanics Ano de publicação: 2014 Tipo de documento: Artigo