Study on radiation dose in different field of views of oral and maxillofacial cone-beam CT / 中华口腔医学杂志
Chinese Journal of Stomatology ; (12): 748-752, 2018.
Artigo em Chinês | WPRIM | ID: wpr-807598
ABSTRACT
Objective@#To analyze the absorptive radiation dose of non-target sensitive organs under the NewTomVGi oral and maxillofacial cone-beam CT scanning in different view fields.@*Methods@#The thermoluminescence dosimeter chip was placed in the non-target sensitive organ of the simulated radiology head model, and 6 scanning fields were measured and analyzed (standard resolution scanning range was 15 cm×15 cm, 15 cm×12 cm, 12 cm×8 cm, 8 cm×8 cm respectively). High definition resolution the radiation absorption dose of thermoluminescence with the scanning range of 12 cm×8 cm, 8 cm×8 cm, respectively.@*Results@#Under the same resolution, the absorptive dose increased with the expansion of scanning field, and the amount of absorbent during high resolution scanning was higher than that during the standard resolution with the same scanning field.@*Conclusions@#New TomVGi oral and maxillofacial cone-beam CT has sighificant difference in radiation dose in different view fields. The appropriate view field should be selected according to the patient′s clinical needs, so as to reduce unnecessary radiation dose during medical imaging examination.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Stomatology Ano de publicação: 2018 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Stomatology Ano de publicação: 2018 Tipo de documento: Artigo