Your browser doesn't support javascript.
loading
Extravascular Migration of a Fractured Inferior Vena Cava Filter Strut
The Korean Journal of Thoracic and Cardiovascular Surgery ; : 224-227, 2017.
Artigo em Inglês | WPRIM | ID: wpr-84707
ABSTRACT
A 20-year-old man presented with a femur fracture and epidural hemorrhage (EDH) following a fall. One month after fracture surgery, swelling developed in both legs, and he was diagnosed as having a deep-vein thrombosis and pulmonary embolism. A retrievable inferior vena cava filter (IVCF) was inserted, because EDH is a contraindication to anticoagulants. Four months later, he complained of abdominal pain, and a computed tomography scan showed a fractured IVCF strut. After percutaneous removal failed 3 times, the IVCF was surgically removed by orthopedists using a portable image intensifier without cardiopulmonary bypass.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Embolia Pulmonar / Veia Cava Inferior / Ponte Cardiopulmonar / Dor Abdominal / Filtros de Veia Cava / Trombose Venosa / Fêmur / Hemorragia / Perna (Membro) / Anticoagulantes Limite: Humanos Idioma: Inglês Revista: The Korean Journal of Thoracic and Cardiovascular Surgery Ano de publicação: 2017 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Embolia Pulmonar / Veia Cava Inferior / Ponte Cardiopulmonar / Dor Abdominal / Filtros de Veia Cava / Trombose Venosa / Fêmur / Hemorragia / Perna (Membro) / Anticoagulantes Limite: Humanos Idioma: Inglês Revista: The Korean Journal of Thoracic and Cardiovascular Surgery Ano de publicação: 2017 Tipo de documento: Artigo