Your browser doesn't support javascript.
loading
A study on the determination method of iodine content in urine by enzyme-lobeled instrument / 中华地方病学杂志
Chinese Journal of Endemiology ; (12): 221-224, 2020.
Artigo em Chinês | WPRIM | ID: wpr-866093
ABSTRACT

Objective:

To establish an enzyme-labeled instrument automatic colorimetric method for determination of iodine content in human urine.

Methods:

The enzyme-labeled instrument automatic colorimetric method was used to determine the iodine content in human urine. The linear range, detection limit and precision of the method were verified. And the urine iodine test results were compared with the results tested by national health industry standard as arsenic-cerium catalytic spectrophotometry.

Results:

The linear range of iodine standard curve of the enzyme-labeled instrument automatic colorimetric method was 0 - 300 μg/L, the linear correlation coefficient ( r) was - 0.999 5 to - 0.999 2, and the detection limit was 6.5 μg/L. The relative standard deviation ( RSD) of urine samples with low, medium and high iodine concentration were all < 3%, the recovery rate ranged from 92.2% to 109.2%, and the total average recovery rate was 99.6%. There was no significant difference in the detection results of iodine content in 40 urine samples between the enzyme-labeled instrument automatic colorimetric method and arsenic-cerium catalytic spectrophotometry ( t = 1.347, P > 0.05); and the detection speed of the enzyme-labeled instrument automatic colorimetric method was 7.5 times of the arsenic-cerium catalytic spectrophotometry.

Conclusion:

The enzyme-labeled instrument automatic colorimetric method has a reasonable linear range, good precision and high accuracy in determination of urinary iodine content, and the enzyme-labeled instrument automatied fast colorimetry has improved the analysis speed, it is suitable for detection of large quantities of samples.
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Endemiology Ano de publicação: 2020 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Endemiology Ano de publicação: 2020 Tipo de documento: Artigo