Research of TLD calibrated in different standard X-ray RQR radiation field in terms of eye lens dose Hp (3) / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 42-46, 2020.
Artigo em Chinês | WPRIM | ID: wpr-868397
ABSTRACT
Objective To compare the calibration result of standard X-ray RQR radiation field between SSDL (NIRP) and CEA LIST LNHB (France),and to explore the feasibility of calibrating Hp (3) in standard X-ray RQR radiation field of SSDL (NIRP).Methods Using a column model with a diameter and high of 20 cm,TLD was calibrated in SSDL (NIRP) and CEA LIST LNHB (France) to measure the personal dose equivalent eye lens dose Hp (3),X-ray RQR radiation field included RQR4 (60 kV),RQR7 (90 kV),RQR9 (120 kV),with energy response,angle response and linear response.Results In terms of energy response,the calibration results of TLD in both SSDL (NIRP) and CEA LIST LNHB (France) were in good agreement.The difference between exposure value and response value was less than 10%.In terms of angle response,the calibration result of TLD in CEA LIST LNHB (France) was better in SSDL (NIRP).The difference between exposure value and response value in CEA LIST LNHB (France) was less than 6%,while the difference between exposure value and response value in SSDI (NIRP) was more than 10% at angle of 20°.In terms of linear response,both calibration result of SSDL (NIRP) and CEA LIST LNHB (France) were in good agreement.Conclusions The standard X-ray RQR field in SSDL (NIRP) can be used for the calibration of Hp (3).

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiological Medicine and Protection Ano de publicação: 2020 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiological Medicine and Protection Ano de publicação: 2020 Tipo de documento: Artigo