OBJECTIVE:

To evaluate the capacity of a computer-aided detection (CAD) system to detect lung nodules in clinical chest CT. MATERIALS AND

METHODS:

A total of 210 consecutive clinical chest CT scans and their reports were reviewed by two chest radiologists and 70 were selected (33 without nodules and 37 with 1-6 nodules, 4-15.4 mm in diameter). The CAD system (ImageChecker (R) CT LN-1000) developed by R2 Technology, Inc. (Sunnyvale, CA) was used. Its algorithm was designed to detect nodules with a diameter of 4-20 mm. The two chest radiologists working with the CAD system detected a total of 78 nodules. These 78 nodules form the database for this study. Four independent observers interpreted the studies with and without the CAD system.

RESULTS:

The detection rates of the four independent observers without CAD were 81% (63/78), 85% (66/78), 83% (65/78), and 83% (65/78), respectively. With CAD their rates were 87% (68/78), 85% (66/78), 86% (67/78), and 85% (66/78), respectively. The differences between these two sets of detection rates did not reach statistical significance. In addition, CAD detected eight nodules that were not mentioned in the original clinical radiology reports. The CAD system produced 1.56 false-positive nodules per CT study. The four test observers had 0, 0.1, 0.17, and 0.26 false-positive results per study without CAD and 0.07, 0.2, 0.23, and 0.39 with CAD, respectively.

CONCLUSION:

The CAD system can assist radiologists in detecting pulmonary nodules in chest CT, but with a potential increase in their false positive rates. Technological improvements to the system could increase the sensitivity and specificity for the detection of pulmonary nodules and reduce these false-positive results.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Radiografia Torácica / Tomografia Computadorizada por Raios X / Diagnóstico por Computador / Sensibilidade e Especificidade / Reações Falso-Positivas / Pneumopatias / Neoplasias Pulmonares Tipo de estudo: Estudo diagnóstico Limite: Humanos Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2005 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Radiografia Torácica / Tomografia Computadorizada por Raios X / Diagnóstico por Computador / Sensibilidade e Especificidade / Reações Falso-Positivas / Pneumopatias / Neoplasias Pulmonares Tipo de estudo: Estudo diagnóstico Limite: Humanos Idioma: Inglês Revista: Korean Journal of Radiology Ano de publicação: 2005 Tipo de documento: Artigo