Your browser doesn't support javascript.
loading
Detection of chloroprene in human blood by headspace gas chromatography / 中国职业医学
China Occupational Medicine ; (6): 722-727, 2016.
Artigo em Chinês | WPRIM | ID: wpr-877017
ABSTRACT

OBJECTIVE:

To establish a detection method for chloroprene in human blood by headspace gas chromatography.

METHODS:

The blood sample was placed directly in the headspace bottle,after the head space automatic sampling,the target object was detected by gas chromatography with electron capture detector,quantified by external standard method.

RESULTS:

The good linearity range of chloroprene was 0. 00-160. 00 μg / L,and the correlation coefficient was 0. 999 4. The detection limit was 0. 10 μg / L and the minimum detectable concentration in blood samples was 0. 10 μg / L. The with-run relative standard deviation( RSD) was 1. 38%-1. 91%,and the between-run RSD was 2. 11%-3. 08%. The average recovery rate was 89. 00%-93. 88%.

CONCLUSION:

The method has the features of high sensitivity,wide linear range and simple sample processing,which was suitable for efficient detection for large quantity of samples.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: China Occupational Medicine Ano de publicação: 2016 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Tipo de estudo: Estudo diagnóstico Idioma: Chinês Revista: China Occupational Medicine Ano de publicação: 2016 Tipo de documento: Artigo