Design of Pulse Signal Acquisition System of Adaptive Wide Dynamic Range Based on ADS125H02 / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 284-287, 2021.
Artigo em Chinês | WPRIM | ID: wpr-880468
ABSTRACT
In order to solve the problem of wide input range of photocurrent, aiming at the shortage of existing pulse signal acquisition system, an adaptive pulse signal acquisition system based on ADS125H02 is designed, which has high collection accuracy, low noise, anti-interference, wide input range, high integration, etc. The measurement range of the system can reach ±10 V, and the accuracy is 0.009 μV. Experimental tests show that the design of the system can extract weak pulse signals, and has good practicality.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Frequência Cardíaca Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2021 Tipo de documento: Artigo