Your browser doesn't support javascript.
loading
Detecting 1,2-dichloropropane and 1,3-dichloropropane in workplace air by solvent desorption-gas chromatography / 中国职业医学
China Occupational Medicine ; (6): 369-372, 2018.
Artigo em Chinês | WPRIM | ID: wpr-881711
ABSTRACT

OBJECTIVE:

To establish a method for simultaneous determination of 1,2-dichloropropane and 1,3-dichloropropane in workplace air by solvent desorption-gas chromatography.

METHODS:

The air samples were collected by activated carbon tube and desorbed with carbon disulfide,then separated by DB-1 capillary columns,and detected by gas chromatographic.

RESULTS:

The linear ranges of 1,2-dichloropropane and 1,3-dichloropropane were 1.16-17 400.00 mg/L and 1.19-17 850.00 mg/L of mass concentration,respectively.The correlation coefficient was 0.999 9,and the limit of detection was 0.07 mg/L.The average desorption efficiencies of 1,2-dichloropropane and 1,3-dichloropropane were 96.9%-98.2%and 95.8%-96.8%,respectively.The within-run relative standard deviation( RSD) was 0.4%-0.6%.The between-run RSD were 1.6%-3.2% and 1.2%-2.8% respectively.The samples could be stored at room temperature for at least 14 days.

CONCLUSION:

This method has high desorption efficiency and sensitivity with good precision and simple operation,which is suitable for simultaneous detection of 1,2-dichloropropane and 1,3-dichloropropane in the workplace air.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: China Occupational Medicine Ano de publicação: 2018 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: China Occupational Medicine Ano de publicação: 2018 Tipo de documento: Artigo