Pull-off resistance of a screwless implantabutment connection and surface evaluation after cyclic loading
The Journal of Advanced Prosthodontics
; : 152-159, 2021.
Article
em En
| WPRIM
| ID: wpr-895774
Biblioteca responsável:
WPRO
ABSTRACT
Purpose@#The aim of this study was to investigate to what extent cyclic load affects the screwless implant-abutment connection for Morse taper dental implants. @*Materials and Methods@#16 implants (SICvantage max) and 16 abutments (Swiss Cross) were used. The screwless implant-abutment connection was subjected to 10,000 cycles of axial loading with a maximum force of 120 N. For the pull-off testing, before and after the same cyclic loading, the required force for disconnecting the remaining 6 implant-abutment connections was measured. The surface of 10 abutments was examined using a scanning electron microscope 120× before and after loading. @*Results@#The pull-off test showed a significant decrease in the vertical force required to pull the abutment from the implant with mean 229.39 N ± 18.23 before loading, and 204.30 N ± 13.51 after loading (P <.01). Apart from the appearance of polished surface areas and slight signs of wear, no visible damages were found on the abutments. @*Conclusion@#The deformation on the polished abutment surface might represent the result of micro movements within the implant-abutment connection during loading. Although there was a decrease of the pull-off force values after cyclic loading, this might not have a notable effect on the clinical performance.