Your browser doesn't support javascript.
loading
Metal Failure of Pedicle Screw System / 대한척추외과학회지
Journal of Korean Society of Spine Surgery ; : 157-163, 2002.
Artigo em Coreano | WPRIM | ID: wpr-92538
ABSTRACT
STUDY

DESIGN:

The metal failure of pedicle screw system followed by posterolateral or posterior fusion used in variable cases(spine fracture, degenerative disorder of spine, deformity of spine) was analyzed retrospectively.

PURPOSE:

The goal of this study was to analyze frequency and clinical consequence of pedicle screw fixation system failure in the treatment of different etiology(spine fracture, degenerative disorder, deformity of spine) and to evaluate affected factors in metal failure. MATERIALS AND

METHODS:

We performed survivorship analysis on 442 patients treated with pedicle screw system from September 1990 to December 1999. The average follow-up period was 54 months(from 18 months to 129 months). As affected factors, some variables such as etiology, kinds of system and extent of fusion were subjected to analyzed their influence on metal failure. We also performed analysis about relationship between metal failure and clinical results. We defined the metal failure as 1) breakage of screw or rod 2) screw bending above 5 degrees 3) dissociation of rod-screw coupling system and 4) screw pull out from vertebral body or pedicle.

RESULTS:

We found out 33 cases of metal failure among 2786 screws, 41 screws had a problem. The metal failure rate was different between each etiology ; 12 cases in fracture(10%), 21 cases in degenerative disorder(6.9%). There was also difference between a kinds of implants; 13 cases in side assembling type(5.4%), 20 cases in back open type(10.5%). However, there were no difference according to extent of fusion; 6 cases in one segment(6.9%), 21 cases in two segments(8.1%), 6 cases in more than three segments(6.3%). Among the overall patients with metal failure (33 cases), only eight patients were complaint significant symptoms. And three of this eight patients were improved after reoperation The mean interval to metal failure was 14.4 months from operation.

CONCLUSION:

The metal failure was more common in spine fracture(p0.05). And metal failure did not significantly affect the clinical results(p>0.05).
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Reoperação / Coluna Vertebral / Anormalidades Congênitas / Taxa de Sobrevida / Estudos Retrospectivos / Seguimentos Tipo de estudo: Estudo observacional / Estudo prognóstico Limite: Humanos Idioma: Coreano Revista: Journal of Korean Society of Spine Surgery Ano de publicação: 2002 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Reoperação / Coluna Vertebral / Anormalidades Congênitas / Taxa de Sobrevida / Estudos Retrospectivos / Seguimentos Tipo de estudo: Estudo observacional / Estudo prognóstico Limite: Humanos Idioma: Coreano Revista: Journal of Korean Society of Spine Surgery Ano de publicação: 2002 Tipo de documento: Artigo