Study on Impedance of Implantable Cardiac Pacemaker in Unipolar/Bipolar Pacing Mode by in Vitro Experiment / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 237-241, 2022.
Artigo em Chinês | WPRIM | ID: wpr-928896
ABSTRACT
The unipolar/bipolar pacing mode of pacemaker is related to its circuit impedance, which affects the battery life. In this study, the in vitro experiment scheme of pacemaker circuit impedance test was constructed. The human blood environment was simulated by NaCl solution, and the experimental environment temperature was controlled by water bath. The results of in vitro experiments showed that under the experimental conditions similar to clinical human parameters, the difference between the circuit impedance of bipolar mode and unipolar mode is 120~200 Ω. The results of the in vitro experiment confirmed that the circuit impedance of bipolar circuit was larger than that of unipolar mode, which was found in clinical practice. The results of this study have reference value to the optimization of pacing mode and the reduction of pacemaker power consumption.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Marca-Passo Artificial / Próteses e Implantes / Estimulação Cardíaca Artificial / Impedância Elétrica Limite: Humanos Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2022 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Marca-Passo Artificial / Próteses e Implantes / Estimulação Cardíaca Artificial / Impedância Elétrica Limite: Humanos Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2022 Tipo de documento: Artigo