ABSTRACT

Objective:

To evaluate the five-year safety and efficacy of the second generation biodegradable polymer sirolimus-eluting stent (EXCROSSAL) in treating patients with de novo coronary artery diseases.

Methods:

Patients with coronary artery disease (CAD)who were implanted with EXTROSSAL stents in CREDIT Ⅱ and CREDIT Ⅲ study were included. CREDIT Ⅱ was a randomized trial, and CREDIT Ⅲ was a single-arm study. From November 2013 to December 2014, 833 CAD patients with de novo coronary lesions implanted with EXTROSSAL stents were selected from 33 centers in China. The primary outcome was 5-year target lesion failure (TLF), a composite of cardiac death, target vessel myocardial infarction and clinically indicated target lesion revascularization. Secondary endpoints was patient-oriented composite endpoint (PoCE), including all-cause death, all myocardial infarction, or any revascularization within 5 years post stenting and stent thrombosis according to Academic Research Consortium's (ARC) definition. Kaplan Meier method was used to calculate the incidence of TLF and PoCE within 5 years after operation. Univariate Cox regression analysis was used to analyze the impacts of diabetes, small vessel disease (vessel diameter ≤ 2.74 mm), lesion length ≥ 16.7 mm and multivessel disease on the incidence of TLF within 5 years after operation.

Results:

A total of 833 patients were included in this study including 579 males (69.5%), the age was (59.3±9.1) years. And 832 (99.9%) patients completed 5-year clinical follow-up. The incidence of TLF and PoCE in the 5-year follow-up were 10.6%(86/811) and 15.5%(126/811), respectively. Stent thrombosis occurred in 1.0%(8/811) of patients. Univariate Cox regression analysis showed that vessel diameter ≤ 2.74 mm (HR=3.20,95%CI 1.90-5.39,P<0.001), lesion length ≥ 16.7 mm (HR=1.88,95%CI 1.18-2.99,P=0.007) and multivessel disease (HR=2.44,95%CI 1.60-3.72,P<0.001) were related factors of TLF within 5 years after operation.

Conclusion:

EXCROSSAL stent is effective and safe in treating CAD patients with de novo coronary lesions, with low incidence of TLF and PoCE within 5 years after operation.
Assuntos
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Polímeros / Fatores de Tempo / Doença da Artéria Coronariana / Fármacos Cardiovasculares / China / Fatores de Risco / Resultado do Tratamento / Sirolimo / Stents Farmacológicos / Intervenção Coronária Percutânea Tipo de estudo: Ensaio Clínico Controlado / Estudo de etiologia / Fatores de risco Limite: Idoso / Humanos / Masculino País/Região como assunto: Ásia Idioma: Chinês Revista: Chinese Journal of Cardiology Ano de publicação: 2021 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Polímeros / Fatores de Tempo / Doença da Artéria Coronariana / Fármacos Cardiovasculares / China / Fatores de Risco / Resultado do Tratamento / Sirolimo / Stents Farmacológicos / Intervenção Coronária Percutânea Tipo de estudo: Ensaio Clínico Controlado / Estudo de etiologia / Fatores de risco Limite: Idoso / Humanos / Masculino País/Região como assunto: Ásia Idioma: Chinês Revista: Chinese Journal of Cardiology Ano de publicação: 2021 Tipo de documento: Artigo