Research on Mathematical Distribution of Failure Time and Engineering Evaluation for Medical Electrical Equipment Based on Operation and Maintenance Data / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 199-204, 2020.
Artigo em Chinês | WPRIM | ID: wpr-942727
ABSTRACT
This study mainly discusses the contents, methods and characteristics of the collection of operation and maintenance data, as well as the establishment and evaluation methods of the distribution model of the failure time of medical electrical equipment. The distribution models of failure time at three levels of medical electrical equipment are established by linear regression method and goodness of fit testThe first is the device level MTBF distribution model, the second is the failure rate distribution model of the failure mode of key components, the third is the calculation model of the influence coefficient of influence factor on the failure mode of key components. This study presents a method of establishing MTBF segment model and implements a calculation model of influence coefficient varying with time.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Manutenção Tipo de estudo: Estudo prognóstico Idioma: Chinês Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2020 Tipo de documento: Artigo