ABSTRACT

Objective:

To explore the accuracy and repeatability of the smartphone scoliosis screening APP developed by our team in the measurement of the Cobb angle of adolescent idiopathic scoliosis (AIS).

Methods:

Clinical data of 60 patients with AIS admitted to Department of Spine and Spinal Surgery, Henan Province People′s Hospital from August 2020 to February 2021 were analyzed retrospectively.Three surveyors measured the coronal main curvature Cobb angle, sagittal thoracic kyphosis (TK) angle, thoracolumbar kyphosis (TLK) angle and lumbar kyphosis (LL) angle on whole-spine lateral X-ray films of 60 patients with AIS by means of the protractor, scoliosis screening APP and computerized Picture Archiving and Communication System (PACS). The time and results of each measurement were recorded.The measurement was repeated once after 2 weeks.Paired t-test was used to compare the measurement time of the APP method and the protractor method.Taking the Cobb angle measurement results of the PACS system as the reference standard, the accuracy of Cobb angle measurement by the APP method was analyzed by paired t-test.The repeatability of the surveyor and the consistency between the surveyors was compared by intraclass correlation coefficient (ICC).

Results:

Among the 60 patients with AIS, there were 17 males and 43 females, aged from 10 to 16 years [(12.2±2.4) years]. The main bends were thoracic curvature (Lenke Ⅰ) in 23 cases, Lenke Ⅱ in 18 cases and thoracolumbar curvature/lumbar curvature (Lenke V) in 19 cases.The APP method took significantly less time to measure the Cobb angle than the protractor method ( P<0.05). There was no significant difference in the Cobb angle measured by the APP method and PACS method ( P>0.05). The results of the coronal main curvature Cobb angle, TK angle, TLK angle and LL angle measured by 3 surveyors through the APP method were all in good agreement (ICC=0.990, 0.988, 0.986, 0.987). The repeatability (ICC 0.973-0.982) of the coronal main curvature Cobb angle, TK, TLK and LL measured twice before and after the APP method were both better that of the protractor method (ICC 0.933-0.954).

Conclusions:

Compared with the traditional protractor, the smartphone scoliosis screening APP has the advantages of short measurement time, high efficiency, excellent accuracy and good repeatability in measuring the Cobb angle of AIS.