Determination of arsenic in urine by hydride generation atomic fluorescence spectrometry with ammonium persulfate digestion / 中华地方病学杂志
Chinese Journal of Endemiology ; (12): 585-590, 2022.
Article em Zh | WPRIM | ID: wpr-955752
Biblioteca responsável: WPRO
ABSTRACT
Objective:To establish a hydride generation atomic fluorescence method using ammonium persulfate as the digestion reagent for determination of arsenic in urine (hereinafter referred to as this method).Methods:The collected urine samples with ammonium persulfate were heated and digested on the tubular electric heating automatic control constant temperature digester (60 holes), with 5% hydrochloric acid solution as reaction medium and current carrier and 1.5% potassium borohydride solution as reducing agent. Arsenic content was determined with a four-channel atomic fluorescence spectrometer. The arsenic standard solution of 0 - 10 μg/L was prepared to determine the standard curve of this method, and the method was evaluated from the detection limit, linear range, correlation coefficient, precision, standard addition recovery experiment, and urine arsenic quality control sample detection. The standard method "Determination of Arsenic in Urine by Hydride Generation Atomic Fluorescence Spectrometry" (WS/T 474-2015, referred to as the standard method) was used for comparison experiments.Results:When the sampling volume was 1 ml, the detection limit of this method (digest with 1 ml 1.5 mol/L ammonium persulfate) was 0.03 μg/L. In the range of arsenic content from 0 - 10 μg/L, the linear relationship between arsenic content and fluorescence intensity was good, and the correlation coefficients ( r) were all 0.999 9. The relative standard deviations( RSD) of the three replicates of urine samples with different concentrations were 1.00%, 0.89% and 0.49%, respectively. Urine arsenic quality control samples were tested, and the test results were all within the range of public values; the overall average recovery was 102.29%, and the recovery range was 92.10% - 108.15%. Compared with the standard method in the determination results of 20 urine samples, the difference was not statistically significant ( t = - 0.40, P > 0.05). Conclusions:The hydride generation atomic fluorescence spectrometry using ammonium persulfate as digestion reagent for the determination of arsenic in urine has the advantages of low detection limit, good precision, high accuracy, small amount of sampling and digestion reagent, simple operation, and less harmful gas generation in sample pretreatment. It is suitable for rapid determination of arsenic in urine in large quantities.
Palavras-chave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: Chinese Journal of Endemiology Ano de publicação: 2022 Tipo de documento: Article
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: Chinese Journal of Endemiology Ano de publicação: 2022 Tipo de documento: Article