ABSTRACT
With the increased use of implantable cardiac devices, the incidence of hardware problems has also increased. Some of the hardware problems might be induced by patient factors. We experienced recurrent pacemaker lead fracture and dislocation after permanent pacemaker implantation. The patient was a bus driver who used his left arm vigorously when he turned the steering wheel. After a new lead was inserted via an axillary vein approach and the patient changed his occupation, no more lead problems have occurred. When a recurrent hardware problem with an implantable pacemaker or defibrillator occurs, patient factors related to anatomy or behavior must be considered.
Assuntos

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Braço / Veia Axilar / Incidência / Desfibriladores / Luxações Articulares / Bloqueio Atrioventricular / Ocupações Tipo de estudo: Estudo de incidência / Estudo prognóstico Limite: Humanos Idioma: Coreano Revista: Korean Journal of Medicine Ano de publicação: 2010 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Assunto principal: Braço / Veia Axilar / Incidência / Desfibriladores / Luxações Articulares / Bloqueio Atrioventricular / Ocupações Tipo de estudo: Estudo de incidência / Estudo prognóstico Limite: Humanos Idioma: Coreano Revista: Korean Journal of Medicine Ano de publicação: 2010 Tipo de documento: Artigo