Objective To establish a detection method for the determination of tetrodotoxin (TTX) in sustained-release microspheres. Methods The HPLC separation of tetrodotoxin was performed on an Agilent ZORBAX SB-C18 column (4.6mm×150mm,5 μm) with acetonitrile, 8mmol/L sodium heptane sulfonate containing 0.005% TFA (595) (pH 4.0) as the mobile phase. The flow rate was 1.0 ml/min. The UV detection wavelength was 200 nm and the column temperature was 30 °C. Results The method had good specificity and linearity of TTX in the concentration range of 1−20 μg/ml. The intra-day precision, inter-day precision, stability and repeatability of the method were good, and the average recoveries were found between 98.0% and 102.0%. Conclusion This study established an HPLC method which was suitable for the determination of tetrodotoxin sustained-release microspheres. The method is accurate and reliable within the applicable range, with strong specificity, which could lead to quantitative detection.