The Reproducibility and Accuracy of Biometry Parameter Measurement from IOL Master(R)
Journal of the Korean Ophthalmological Society ; : 1665-1673, 2004.
Artigo em Coreano | WPRIM | ID: wpr-97117
ABSTRACT

PURPOSE:

To evaluate the reproducibility and accuracy of axial length and anterior chamber depth measurements from IOL Master(R).

METHODS:

The axial length and anterior chamber depth measurements in 30 eyes with normal crystalline lens, 30 eyes with cataract and 30 pseudophakic eyes were measured by two practitioners using IOL Master(R) followed by A-scan and Orbscan II. The reproducibility of IOL Master(R) was analyzed by comparing the results from the first and second practitioners. To evaluate the accuracy of IOL Master(R), the axial length was compared to A-scan and the anterior chamber depth was compared to A-scan and Orbscan II.

RESULTS:

The difference between the axial length and anterior chamber depth measurements from the two practitioners using IOL Master(R) were not statistically significant (p>0.05). The axial length of measurement from IOL Master(R) was 0.16 mm, 0.18 mm, and 0.96 mm longer than that from A-scan, in the normal, cataract and pseudophakic groups, respectively. In addition, the anterior chamber depth measurement from IOL Master(R) was significantly deeper than that from A-scan and Orbscan II (p<0.05).

CONCLUSIONS:

There was good reproducibility and accuracy of axial length and anterior chamber depth measurements from IOL Master(R). However, some patients who had media opacity were not measured using IOL Master(R). It is a noncontact method, which provides an alternative to A-scan.