Your browser doesn't support javascript.
loading
GE Signa HDxt 1.5 T MRI PAC Module Troubleshooting and Repair / 中国医疗器械杂志
Article em Zh | WPRIM | ID: wpr-982267
Biblioteca responsável: WPRO
ABSTRACT
Through the analysis of PAC module faults in two GE Signa HDxt 1.5 T MRI cases, the fault performance was analyzed. According to the working principle of PAC module, the fault was analyzed and processed, and then the suspicious fault parts were located and repaired to make the equipment run normally. Finally, the operating principles of PAC module in terms of power supply, signal transmission and data transmission were summarized to facilitate the subsequent quickly and find out fault points for maintenance accurately.
Palavras-chave
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2023 Tipo de documento: Article
Texto completo: 1 Índice: WPRIM Idioma: Zh Revista: Chinese Journal of Medical Instrumentation Ano de publicação: 2023 Tipo de documento: Article