Cochlear electrode array misplacement into the superior semicircular canal: a case report and literature review / 临床耳鼻咽喉头颈外科杂志
Journal of Clinical Otorhinolaryngology Head and Neck Surgery
;
(12): 310-312, 2023.
Artigo
em Chinês
| WPRIM
| ID: wpr-982739
ABSTRACT
Electrode array misplacement is a rare complication of cochlear implant. This article reports an 11-year-old boy who was mistakenly implanted the cochlear electrode array into the superior semicircular canal during the initial cochlear implant. After the diagnosis was confirmed, he underwent a second cochlear implant and the electrode array were successfully implanted into the cochlea. This article conducted a systematic review of the literature on electrode array misplacement, and the causes of electrode array misplacement were analyzed from different implantation position.