Your browser doesn't support javascript.
loading
Evaluation of technical performance of stereotactic radiosurgery algorithm in optical surface imaging system in non-coplanar radiotherapy / 中华放射肿瘤学杂志
Chinese Journal of Radiation Oncology ; (6): 438-444, 2023.
Artigo em Chinês | WPRIM | ID: wpr-993211
ABSTRACT

Objective:

To evaluate the accuracy and stability of stereotactic radiosurgery (SRS) algorithm in optical surface imaging (OSI) system in non-coplanar radiotherapy.

Methods:

Three OSI imaging systems were used to measure the phantom repeatedly at different couch rotation angles to analyze the accuracy and stability of OSI system. Seven patients with multiple brain metastases who underwent single-center non-coplanar radiotherapy were randomly selected, and the accuracy and stability of OSI for patient imaging were analyzed. Stability is defined as the difference between the two OSI measurements when the couch is turned from 0° to a non 0° angle, and then back to 0°, using the 0° cone beam CT (CBCT) as the "gold standard". Accuracy is defined as the difference between OSI and CBCT (at 0° couch angle) measurement data. The measurement data with normal distribution were described as Mean ± SD. The data with non-normal distribution were expressed as M (Q). The difference of the former data was compared by one-way ANOVA, and the difference of the latter data was assessed by Kruskal-Wallis H nonparametric test.

Results:

For non-coplanarity, the translation accuracy of the phantom and the patient was ≤ 1.30 mm and ≤ 1.00 mm, and the rotation accuracy was ≤ 0.50° and ≤ 0.60°, respectively. The translation errors mainly occurred in the left-right and head-foot directions. In terms of stability, the maximum standard deviation of phantom coplanar translation and rotation was 0.06 mm and 0.06°. The maximum standard deviation of patient translation and rotation was 0.17 mm and 0.19°.

Conclusions:

Although the new SRS algorithm improves the non-coplanar accuracy, it still cannot meet the precise requirements of non-coplanar single isocenter radiotherapy for multiple brain metastases, especially in the left-right and head-foot directions. When the couch rotation angle is large, OSI is not recommended for image-guided radiotherapy. However, its high stability can be used to monitor the intrafractional motion of patients.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiation Oncology Ano de publicação: 2023 Tipo de documento: Artigo

Similares

MEDLINE

...
LILACS

LIS

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Radiation Oncology Ano de publicação: 2023 Tipo de documento: Artigo