ABSTRACT

Objective:

To investigate the significance and importance of the interaction between surgeons and engineers during the preoperative planning phase of total knee arthroplasty (TKA) when utilizing patient-specific instrumentation (PSI).

Methods:

A retrospective review was conducted on 202 knees of PSI-assisted TKA performed on 178 patients between June 2018 and August 2022. The patients' mean age was 68.4±6.2 years, ranging from 53 to 86 years. Among the participants, there were 149 females and 29 males, 93 left knees and 109 right knees. The study involved 171 patients of osteoarthritis (193 knees) and 7 patients of rheumatoid arthritis (9 knees), with 194 knees presenting varus knees and 8 knees with valgus knees. The preoperative plan documents, from the initial engineer-designed plan to the final plan approved by the surgeon, were analyzed to assess the frequency, parameters, and reasons for adjustments made during the planning process.

Results:

The planning of the 202 PSI-assisted TKA was subjected to at least one round of surgeon-engineer interaction. Among the 202 TKA planning, 117 knees (57.9%) underwent modifications after discussion, with most plans (100 knees, 49.5%) being confirmed after one round of modification. Two rounds of modifications were performed on 10 knees (5.0%), and three rounds on 5 knees (2.5%). A maximum of four rounds of modifications were made on two knees (0.9%). Furthermore, in the case of the remaining 85 knees (42.1%), the surgeons promptly consented to the engineers' initial planning following the discussions. Specific adjustments were made in 106 knees (52.5%) regarding femoral parameters, 57 knees (28.2%) concerning tibial parameters, and 46 knees (22.8%) requiring adjustments to both femoral and tibial parameters. Notably, the most frequently adjusted parameter was the osteotomy thickness of the posterior femoral condyles, which was modified in 94 knees (80.3%). The reasons for adjusting femoral or tibial parameters were summarized, revealing the main factors as follows 1) Discrepancy between the mediolateral and anteroposterior diameters of the femoral condyle; 2) Twisted deformity of the proximal tibia; 3) Severe flexion contracture deformity of the knee; 4) Collapse of the medial or lateral tibial plateau; 5) Evident anterior arch deformity of the femur.

Conclusion:

The interaction between surgeons and engineers plays a pivotal role in the preoperative phase of PSI-assisted TKA. Effective collaboration allows surgeons to accurately analyze the unique anatomical characteristics and pathological changes of each patient in a three-dimensional perspective, facilitating the formulation of individualized surgical plans.