Your browser doesn't support javascript.
loading
A prospective multicenter clinical trial study of a domestic HeartCon third-generation magnetic and hydrodynamic levitation LVAD for the treatment of 50 cases of end-stage heart failure / 中华胸心血管外科杂志
Chinese Journal of Thoracic and Cardiovascular Surgery ; (12): 273-278, 2023.
Artigo em Chinês | WPRIM | ID: wpr-995551
ABSTRACT

Objective:

To analyze and evaluate the safety and efficacy of a Chinese domestically manufactured Heart Con-type implantable third-generation magnetic and hydrodynamic levitation left ventricular assist device(LVAD) for the treatment of end-stage heart failure(ESHF), by reporting the results of eleven-center clinical trial on 50 cases.

Methods:

This study was a multicenter clinical trial, designed by means of prospective, multicenter and single-group target value. 50 subjects with ESHF were competitively enrolled and treated with HeartCon as the LVAD in eleven centers. The primary efficacy measure was survival, defined as either the subjects experiencing the transition to heart transplantation(HT) or myocardial recovery assisted by the device within 90 days, or as successfully assisted by the LVAD for full 90 days after implantation. The target survival rate was 60%, other observations included implantation success rate, mortality, pump failure needing replacement or emergency heart transplantation.

Results:

All enrolled 50 patients received LVAD implantation successfully, 46 survived with the pump for 90 days, 1 patient transitioned to heart transplantation, and 3 patients experienced pump thrombosis, within which 2 patients underwent pump replacement and continued to live with the pump for 90 days, and the other one received emergency heart transplantation. There were no dropout subjects. The survival rate at full 90 days after HeartCon implantation was 100%. The survival rates with pump in the full set analysis and the protocol set analysis were 96.00% and 95.92% respectively, which were higher than the target value of 60%. The differences were both statistically significant( P<0.05).

Conclusion:

The results of the multicenter clinical trial with the largest sample size in China using domestically manufactured third-generation LVAD has demonstrated that, HeartCon is a safe and effective LVAD to treat ESHF patients.

Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Chinês Revista: Chinese Journal of Thoracic and Cardiovascular Surgery Ano de publicação: 2023 Tipo de documento: Artigo