Your browser doesn't support javascript.
loading
The Feasibility of Multiple Fixation Points in C2
Asian Spine Journal ; : 888-893, 2023.
Artigo em Inglês | WPRIM | ID: wpr-999653
ABSTRACT
Methods@#We used 1.0-mm interval computed tomographic scan images of 100 patients (50 men and 50 women) and screw trajectory simulation software. The diameter of all screws was set at 3.5 mm, considering its common usage in real surgery. The anatomical feasibility of placing both pedicle and laminar screws on the same side was evaluated. For all feasible sides, the three-dimensional distance between the screw entry points was measured. @*Results@#In 85% of cases, both pedicle and laminar screws could be placed on both sides, allowing for the insertion of 4 screws. In 11% of cases, 2 screws could be placed on one side, while only 1 screw was feasible on the other side, resulting in the placement of 3 screws. In all 181 sides where both types of screws could be inserted, the distance between their entry points exceeded 16.1 mm, which was sufficient to prevent the collision between the screw heads. @*Conclusions@#C2 vertebra can accommodate three (11%) or four (85%) screws in 96% of cases.
Texto completo: DisponíveL Índice: WPRIM (Pacífico Ocidental) Idioma: Inglês Revista: Asian Spine Journal Ano de publicação: 2023 Tipo de documento: Artigo

Similares