Your browser doesn't support javascript.
Point-of-care testing for the detection of SARS-CoV-2: a systematic review and meta-analysis.
Yoon, S H; Yang, S; Cho, H; Eun, S; Koo, C M; Kim, M K.
  • Yoon SH; Department of Pediatrics, Severance Children's Hospital, Yonsei University College of Medicine, Seoul, South Korea. YOONSH@yuhs.ac.
Eur Rev Med Pharmacol Sci ; 25(1): 503-517, 2021 Jan.
Article in English | MEDLINE | ID: covidwho-1052577
ABSTRACT

OBJECTIVE:

To evaluate the diagnostic accuracy of the Food and Drug Administration Emergency Use Authorization (FDA-EUA) authorized point-of-care tests (POCTs) for the detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). MATERIALS AND

METHODS:

A systematic literature search was conducted using the PubMed, Embase, and Web of Science databases for articles published till August 10, 2020. We included studies providing information regarding diagnostic test accuracy of FDA-EUA POCTs for SARS-CoV-2 detection. The methodologic quality was assessed using the Quality Assessment of Diagnostic Accuracy Studies-2 tool. The review protocol is registered in the International Prospective Register of Systematic Reviews (protocol number CRD42020202248).

RESULTS:

We included 26 studies describing a total of 3242 samples. The summary sensitivity and specificity were 0.94 [95% confidence interval (CI) 0.88-0.97] and 1.00 (95% CI 0.99-1.00), respectively. The area under the summary receiver operating characteristic curve was 1.00 (95% CI 0.99-1.00). A pooled analysis based on the index test revealed a summary sensitivity and specificity of Cepheid Xpert Xpress SARS-CoV-2 [0.99 (95% CI 0.97-1.00) and 0.99 (95% CI 0.94-1.00, respectively)] and ID NOW COVID-19 [0.78 (95% CI 0.74-0.82) and 1.00 (95% CI 0.98-1.00), respectively].

CONCLUSIONS:

FDA-EUA POCTs, especially molecular assays, have high sensitivity, specificity, and overall diagnostic accuracy for detecting SARS-CoV-2. If approved, FDA-EUA POCTs can provide a rapid and practical way to identify infected individuals early on and help to limit the strain on the healthcare system. However, more high-quality clinical data are required to support our results.
Subject(s)

Full text: Available Collection: International databases Database: MEDLINE Main subject: Point-of-Care Testing / COVID-19 Testing / SARS-CoV-2 / COVID-19 Type of study: Diagnostic study / Experimental Studies / Observational study / Prognostic study / Reviews / Systematic review/Meta Analysis Limits: Humans Country/Region as subject: North America Language: English Journal: Eur Rev Med Pharmacol Sci Journal subject: Pharmacology / Toxicology Year: 2021 Document Type: Article Affiliation country: Eurrev_202101_24422

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Point-of-Care Testing / COVID-19 Testing / SARS-CoV-2 / COVID-19 Type of study: Diagnostic study / Experimental Studies / Observational study / Prognostic study / Reviews / Systematic review/Meta Analysis Limits: Humans Country/Region as subject: North America Language: English Journal: Eur Rev Med Pharmacol Sci Journal subject: Pharmacology / Toxicology Year: 2021 Document Type: Article Affiliation country: Eurrev_202101_24422