Your browser doesn't support javascript.
Diagnostic performance of a SARS-CoV-2 rapid antigen test in a large, Norwegian cohort.
Landaas, Elisabeth Toverud; Storm, Margrethe Larsdatter; Tollånes, Mette Christophersen; Barlinn, Regine; Kran, Anne-Marte Bakken; Bragstad, Karoline; Christensen, Andreas; Andreassen, Trude.
  • Landaas ET; Department of Microbiology, Oslo University Hospital, Oslo, Norway; Institute of Clinical Medicine, University of Oslo, Oslo, Norway. Electronic address: eltlan@ous-hf.no.
  • Storm ML; Division of Infection Control and Environmental Health, Norwegian Institute of Public Health, Oslo, Norway.
  • Tollånes MC; Norwegian Organization for Quality Improvement of Laboratory Examinations (Noklus), Haraldsplass Deaconess Hospital, Bergen, Norway; Department of Global Public Health and Primary Care, University of Bergen, Bergen, Norway.
  • Barlinn R; Department of Microbiology, Oslo University Hospital, Oslo, Norway.
  • Kran AB; Division of Infection Control and Environmental Health, Norwegian Institute of Public Health, Oslo, Norway.
  • Bragstad K; Division of Infection Control and Environmental Health, Norwegian Institute of Public Health, Oslo, Norway.
  • Christensen A; Division of Infection Control and Environmental Health, Norwegian Institute of Public Health, Oslo, Norway; Department of Medical Microbiology, St. Olav's Hospital, Trondheim, Norway.
  • Andreassen T; Norwegian Directorate of Health, Oslo, Norway.
J Clin Virol ; 137: 104789, 2021 04.
Article in English | MEDLINE | ID: covidwho-1126918
ABSTRACT

BACKGROUND:

Rapid antigen tests (RATs) may be included in national strategies for handling the SARS-CoV-2 pandemic, as they provide test results rapidly, are easily performed outside laboratories, and enable immediate contract tracing. However, before implementation further clinical evaluation of test sensitivity is warranted.

OBJECTIVES:

To examine the performance of Abbott's Panbio™ COVID-19 Ag Rapid Test Device for SARS-CoV-2 testing in a low to medium prevalence setting in Norway. STUDY

DESIGN:

A prospective study comparing the results of the Panbio RAT with PCR in 4857 parallel samples collected at a SARS-CoV-2 test station in Oslo, and from COVID-19 outbreaks in six Norwegian municipalities.

RESULTS:

A total of 4857 cases were included in the study; 3991 and 866 cases from the test station and the outbreak municipalities, respectively. The prevalence at the test station in Oslo was 6.3 %, and the overall sensitivity of the RAT was 74 %. Increased sensitivity was observed in patients who experienced symptoms (79 %) and when considering samples with viral loads above estimated level of infectivity (84 %), while it was lower in asymptomatic persons (55 %). In the outbreak municipalities, the overall prevalence was 6.9 %, and the total sensitivity of the RAT was 70 %.

CONCLUSIONS:

Our results indicate that the test correctly identified most infectious individuals. Nevertheless, the sensitivity is considerably lower than for PCR, and it is important that the limitations of the test are kept in mind in the follow-up of tested individuals.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: COVID-19 Serological Testing / SARS-CoV-2 / COVID-19 / Antigens, Viral Type of study: Cohort study / Diagnostic study / Experimental Studies / Observational study / Prognostic study Limits: Humans Country/Region as subject: Europa Language: English Journal: J Clin Virol Journal subject: Virology Year: 2021 Document Type: Article

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: COVID-19 Serological Testing / SARS-CoV-2 / COVID-19 / Antigens, Viral Type of study: Cohort study / Diagnostic study / Experimental Studies / Observational study / Prognostic study Limits: Humans Country/Region as subject: Europa Language: English Journal: J Clin Virol Journal subject: Virology Year: 2021 Document Type: Article