Your browser doesn't support javascript.
Ventilating two patients with one ventilator: technical setup and laboratory testing.
de Jongh, Frans H C; de Vries, Heder J; Warnaar, Rob S P; Oppersma, Eline; Verdaasdonk, Rudolf; Heunks, Leo M A; Doorduin, Jonne.
  • de Jongh FHC; Dept of Pulmonology, Medisch Spectrum Twente, Enschede, The Netherlands.
  • de Vries HJ; TechMed Centre, University of Twente, Enschede, The Netherlands.
  • Warnaar RSP; Dept of Neonatal Intensive Care Medicine, Amsterdam UMC, Amsterdam, The Netherlands.
  • Oppersma E; These authors contributed equally.
  • Verdaasdonk R; Dept of Intensive Care Medicine, Amsterdam UMC, Amsterdam, The Netherlands.
  • Heunks LMA; Amsterdam Cardiovascular Sciences Research Institute, Amsterdam UMC, Amsterdam, The Netherlands.
  • Doorduin J; These authors contributed equally.
ERJ Open Res ; 6(2)2020 Apr.
Article in English | MEDLINE | ID: covidwho-1177705
ABSTRACT
With a modified circuit, it is feasible to ventilate two patients with one ventilator over a relevant range of compliances. Adding inspiratory resistance allows individual titration of tidal volume, and incorporating one-way valves prevents pendelluft. https//bit.ly/3ex8SYP.

Full text: Available Collection: International databases Database: MEDLINE Language: English Year: 2020 Document Type: Article Affiliation country: 23120541.00256-2020

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Language: English Year: 2020 Document Type: Article Affiliation country: 23120541.00256-2020