Your browser doesn't support javascript.
A novel approach to anterior segment imaging with smartphones in the COVID-19 era.
Dutt, Sreetama; Vadivel, Siva S; Nagarajan, Shanmuganathan; Galagali, Amrutha; Christy, Josephine S; Sivaraman, Anand; Rao, Divya Parthasarathy.
  • Dutt S; Research & Development, Remidio Innovative Solutions Pvt. Ltd., Bengaluru, Karnataka, India.
  • Vadivel SS; Research & Development, Remidio Innovative Solutions Pvt. Ltd., Bengaluru, Karnataka, India.
  • Nagarajan S; Research & Development, Remidio Innovative Solutions Pvt. Ltd., Bengaluru, Karnataka, India.
  • Galagali A; Research & Development, Remidio Innovative Solutions Pvt. Ltd., Bengaluru, Karnataka, India.
  • Christy JS; Cornea & Refractive Services, Aravind Eye Hospital, Pondicherry, India.
  • Sivaraman A; Research & Development, Remidio Innovative Solutions Pvt. Ltd., Bengaluru, Karnataka, India.
  • Rao DP; Research & Development, Remidio Innovative Solutions Pvt. Ltd., Bengaluru, Karnataka, India.
Indian J Ophthalmol ; 69(5): 1257-1262, 2021 May.
Article in English | MEDLINE | ID: covidwho-1207864
ABSTRACT

PURPOSE:

To report a novel, telemedicine-friendly, smartphone-based, wireless anterior segment device with instant photo-documentation ability in the COVID-19 era.

METHODS:

Anterior Imaging Module (AIM) was constructed based on a 50/50 beam splitter design, to match the magnification drum optics of slit-lamps with a three-step or higher level of magnification. The design fills the smartphone sensor fully at the lowest magnification and matches the fixed focus of the slit-lamp. It comes with a smartphone for instant photo-documentation, an in-built software application for data-management and secure HIPAA compliant cloud storage, and a Bluetooth trigger for a one-tap image capture. The construction of the device is explained, and the optical resolution measured using U.S. air-force resolution test. AIM's performance was characterized with traceability to internationally relevant performance standards for digital slit-lamps after image quality assessment through a pilot study.

RESULTS:

Clinically useful anterior segment images were obtained with both diffuse and slit illumination at different magnification settings with the highest magnification (40X) resolution of 359 lines per cm and the lowest magnification (16X) resolution of 113 lines per cm.

CONCLUSION:

AIM is a novel, wireless, telemedicine-enabled design that digitizes existing, analog slit lamps with at least three-step magnification. The settings ensure the focus is determined purely by the position of the slit-lamp. Hence, the image viewed and captured on the smartphone is exactly what the clinician sees through the eyepiece. This helps in maintaining distance from the patient in the ongoing COVID-19 pandemic, as well.