Your browser doesn't support javascript.
Performance of the LIAISON® SARS-CoV-2 Antigen Assay vs. SARS-CoV-2-RT-PCR.
Fiedler, Melanie; Holtkamp, Caroline; Dittmer, Ulf; Anastasiou, Olympia E.
  • Fiedler M; Institute for Virology, University Hospital Essen, University of Duisburg-Essen, 45147 Essen, Germany.
  • Holtkamp C; Institute for Virology, University Hospital Essen, University of Duisburg-Essen, 45147 Essen, Germany.
  • Dittmer U; Institute for Virology, University Hospital Essen, University of Duisburg-Essen, 45147 Essen, Germany.
  • Anastasiou OE; Institute for Virology, University Hospital Essen, University of Duisburg-Essen, 45147 Essen, Germany.
Pathogens ; 10(6)2021 May 26.
Article in English | MEDLINE | ID: covidwho-2259131
ABSTRACT
We aimed to evaluate the LIAISON® SARS-CoV-2 antigen assay (DiaSorin), comparing its performance to real-time polymerase chain reaction (RT-PCR) for the detection of SARS-CoV-2 RNA. 182 (110 PCR-positive and 72 PCR-negative) nasopharyngeal swab samples were taken for the detection of SARS-CoV-2. RT-PCR and antigen assay were performed using the same material. The sensitivity and specificity of the antigen assay were calculated for different cut-offs, with RT-PCR serving as the reference method. Stored clinical samples that were positive for other respiratory viruses were tested to evaluate cross-reactivity. One third (33/110, 30%) were falsely classified as negative, while no false positives were found using the 200 TCID50/mL cut-off for the SARS-CoV-2 antigen as proposed by the manufacturer. This corresponded to a sensitivity of 70% (60-78%) and a specificity of 100% (94-100%). Lowering the cut-off for positivity of the antigen assay to 22.79 or 57.68 TCID50/mL increased the sensitivity of the method, reaching a sensitivity of 92% (85-96%) vs. 79% (70-86%) and a specificity of 81% (69-89%) vs. 99% (91-100%), respectively. The antigen assay reliably detected samples with high SARS-CoV-2 viral loads (≥106 copies SARS-CoV-2/mL), while it cannot differentiate between negative and low positive samples. Cross-reactivity toward other respiratory viruses was not detected.
Keywords

Full text: Available Collection: International databases Database: MEDLINE Type of study: Diagnostic study / Experimental Studies / Prognostic study / Randomized controlled trials Language: English Year: 2021 Document Type: Article Affiliation country: Pathogens10060658

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Type of study: Diagnostic study / Experimental Studies / Prognostic study / Randomized controlled trials Language: English Year: 2021 Document Type: Article Affiliation country: Pathogens10060658