Your browser doesn't support javascript.
Use, Safety Assessment, and Implementation of Two Point-of-Care Tests for COVID-19 Testing.
Hahn, Megan; Olsen, Aaron; Stokes, Kindra; Fowler, Randal C; Gu, Rui; Semple-Lytch, Shellanne; DeVito, Andrea; Kurpiel, Philip; Hughes, Scott; Rakeman, Jennifer L.
  • Hahn M; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
  • Olsen A; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
  • Stokes K; Centers for Disease Control and Prevention, Atlanta, GA, USA.
  • Fowler RC; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
  • Gu R; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
  • Semple-Lytch S; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
  • DeVito A; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
  • Kurpiel P; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
  • Hughes S; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
  • Rakeman JL; Department of Health and Mental Hygiene, New York City Public Health Laboratory, New York, NY, USA.
Am J Clin Pathol ; 156(3): 370-380, 2021 08 04.
Article in English | MEDLINE | ID: covidwho-1322609
ABSTRACT

OBJECTIVES:

The Abbot ID NOW COVID-19 assay and Quidel Sofia 2 SARS Antigen FIA are point-of-care assays that offer rapid testing for severe acute respiratory syndrome coronavirus 2 viral RNA and nucleocapsid protein, respectively. Given the utility of these devices in the field, we investigated the feasibility and safety of using the ID NOW and Sofia assays in the public health response to the coronavirus disease 2019 pandemic and in future public health emergencies.

METHODS:

A combination of utilization and contamination testing in addition to a review of instrument workflows was conducted.

RESULTS:

Utilization testing demonstrated that both tests are intuitive, associated with high user test success (85%) in our study, and could be implemented by staff after minimal training. Contamination tests revealed potential biosafety concerns due to the open design of the ID NOW instrument and the transfer mechanisms with the Sofia. When comparing the workflow of the ID NOW and the Sofia, we found that the ID NOW was more user-friendly and that the transfer technology reduces the chance of contamination.

CONCLUSIONS:

The ID NOW, Sofia, and other emerging point-of-care tests should be used only after careful consideration of testing workflow, biosafety risk mitigations, and appropriate staff training.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: Pandemics / Point-of-Care Testing / COVID-19 Testing / SARS-CoV-2 / COVID-19 / Antigens, Viral Type of study: Diagnostic study / Observational study / Prognostic study Limits: Humans Language: English Journal: Am J Clin Pathol Year: 2021 Document Type: Article Affiliation country: Ajcp

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Pandemics / Point-of-Care Testing / COVID-19 Testing / SARS-CoV-2 / COVID-19 / Antigens, Viral Type of study: Diagnostic study / Observational study / Prognostic study Limits: Humans Language: English Journal: Am J Clin Pathol Year: 2021 Document Type: Article Affiliation country: Ajcp