Your browser doesn't support javascript.
Scalable, effective, and rapid decontamination of SARS-CoV-2 contaminated N95 respirators using germicidal ultraviolet C (UVC) irradiation device.
Rathnasinghe, Raveen; Karlicek, Robert F; Schotsaert, Michael; Koffas, Mattheos; Arduini, Brigitte L; Jangra, Sonia; Wang, Bowen; Davis, Jason L; Alnaggar, Mohammed; Costa, Anthony; Vincent, Richard; García-Sastre, Adolfo; Vashishth, Deepak; Balchandani, Priti.
  • Rathnasinghe R; Department of Microbiology, Icahn School of Medicine at Mount Sinai, New York, NY, 10029, USA.
  • Karlicek RF; Graduate School of Biomedical Sciences, Icahn School of Medicine at Mount Sinai, New York, NY, 10029, USA.
  • Schotsaert M; Global Health and Emerging Pathogens Institute, Icahn School of Medicine at Mount Sinai, New York, NY, 10029, USA.
  • Koffas M; Smart Lighting Engineering Research Center, Rensselaer Polytechnic Institute, Troy, NY, 12180, USA.
  • Arduini BL; Department of Electrical Computer and Systems Engineering, Rensselaer Polytechnic Institute, Troy, NY, 12180, USA.
  • Jangra S; Department of Microbiology, Icahn School of Medicine at Mount Sinai, New York, NY, 10029, USA.
  • Wang B; Global Health and Emerging Pathogens Institute, Icahn School of Medicine at Mount Sinai, New York, NY, 10029, USA.
  • Davis JL; Department of Chemical and Biological Engineering, Rensselaer Polytechnic Institute, Troy, NY, 12180, USA.
  • Alnaggar M; Center for Biotechnology and Interdisciplinary Studies, Rensselaer Polytechnic Institute, Troy, NY, 12180, USA.
  • Costa A; Department of Biological Sciences, Rensselaer Polytechnic Institute, Troy, NY, 12180, USA.
  • Vincent R; Center for Biotechnology and Interdisciplinary Studies, Rensselaer Polytechnic Institute, Troy, NY, 12180, USA.
  • García-Sastre A; Department of Microbiology, Icahn School of Medicine at Mount Sinai, New York, NY, 10029, USA.
  • Vashishth D; Center for Biotechnology and Interdisciplinary Studies, Rensselaer Polytechnic Institute, Troy, NY, 12180, USA.
  • Balchandani P; Department of Biomedical Engineering, Rensselaer Polytechnic Institute, Troy, NY, 12180, USA.
Sci Rep ; 11(1): 19970, 2021 10 07.
Article in English | MEDLINE | ID: covidwho-1462030
Preprint
This scientific journal article is probably based on a previously available preprint. It has been identified through a machine matching algorithm, human confirmation is still pending.
See preprint
ABSTRACT
Particulate respirators such as N95s are an essential component of personal protective equipment (PPE) for front-line workers. This study describes a rapid and effective UVC irradiation system that would facilitate the safe re-use of N95 respirators and provides supporting information for deploying UVC for decontamination of SARS-CoV-2 during the COVID-19 pandemic. To assess the inactivation potential of the proposed UVC germicidal device as a function of time by using 3 M 8211-N95 particulate respirators inoculated with SARS-CoV-2. A germicidal UVC device to deliver tailored UVC dose was developed and test coupons (2.5 cm2) of the 3 M-N95 respirator were inoculated with 106 plaque-forming units (PFU) of SARS-CoV-2 and were UV irradiated. Different exposure times were tested (0-164 s) by fixing the distance between the lamp and the test coupon to 15.2 cm while providing an exposure of at least 5.43 mWcm-2. Primary measure of outcome was titration of infectious virus recovered from virus-inoculated respirator test coupons after UVC exposure. Other measures included the method validation of the irradiation protocol, using lentiviruses (biosafety level-2 agent) and establishment of the germicidal UVC exposure protocol. An average of 4.38 × 103 PFU ml-1 (SD 772.68) was recovered from untreated test coupons while 4.44 × 102 PFU ml-1 (SD 203.67), 4.00 × 102 PFU ml-1 (SD 115.47), 1.56 × 102 PFU ml-1 (SD 76.98) and 4.44 × 101 PFU ml-1 (SD 76.98) was recovered in exposures 2, 6, 18 and 54 s per side respectively. The germicidal device output and positioning was monitored and a minimum output of 5.43 mW cm-2 was maintained. Infectious SARS-CoV-2 was not detected by plaque assays (minimal level of detection is 67 PFU ml-1) on N95 respirator test coupons when irradiated for 120 s per side or longer suggesting 3.5 log reduction in 240 s of irradiation, 1.3 J cm-2. A scalable germicidal UVC device to deliver tailored UVC dose for rapid decontamination of SARS-CoV-2 was developed. UVC germicidal irradiation of N95 test coupons inoculated with SARS-CoV-2 for 120 s per side resulted in 3.5 log reduction of virus. These data support the reuse of N95 particle-filtrate apparatus upon irradiation with UVC and supports use of UVC-based decontamination of SARS-CoV-2 during the COVID-19 pandemic.
Subject(s)

Full text: Available Collection: International databases Database: MEDLINE Main subject: Ultraviolet Rays / Decontamination / N95 Respirators / SARS-CoV-2 / COVID-19 Type of study: Experimental Studies / Prognostic study Limits: Animals / Humans Language: English Journal: Sci Rep Year: 2021 Document Type: Article Affiliation country: S41598-021-99431-5

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: Ultraviolet Rays / Decontamination / N95 Respirators / SARS-CoV-2 / COVID-19 Type of study: Experimental Studies / Prognostic study Limits: Animals / Humans Language: English Journal: Sci Rep Year: 2021 Document Type: Article Affiliation country: S41598-021-99431-5