Bain H: Innovative Modification of Bain Circuit for the Resuscitation and Transportation of Patients With Coronavirus Disease 2019.
A A Pract
; 15(10): e01530, 2021 Oct 20.
Article
in English
| MEDLINE | ID: covidwho-1484822
ABSTRACT
Bain H circuit is an innovatively modified breathing circuit designed for the transportation and resuscitation of patients with coronavirus disease (COVID-19). For this circuit, the Heidbrink valve was replaced with a 15F inlet and 15M/22F outlet adjustable pressure-limiting (APL) valve, and a high-efficiency particulate air filter was placed over the APL outlet valve. The circuit is designed to filter the novel coronavirus without any increase in dead space or resistance. All benefits of the conventional Bain circuit were retained. Besides its use in dedicated COVID-19 areas, this circuit can be used in other emergency units of the hospital.
Full text:
Available
Collection:
International databases
Database:
MEDLINE
Main subject:
COVID-19
Limits:
Humans
Language:
English