Field Performance and Public Health Response Using the BinaxNOWTM Rapid Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Antigen Detection Assay During Community-Based Testing.
Clin Infect Dis
; 73(9): e3098-e3101, 2021 11 02.
Article
in English
| MEDLINE | ID: covidwho-1702354
ABSTRACT
Among 3302 persons tested for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) by BinaxNOWTM and reverse transcription polymerase chain reaction (RT-PCR) in a community setting, rapid assay sensitivity was 100%/98.5%/89% using RT-PCR cycle thresholds of 30, 35, and no threshold. The specificity was 99.9%. Performance was high across ages and those with and without symptoms. Rapid resulting permitted immediate public health action.