X-rays through the looking glass: Mobile imaging dosimetry and image quality of suspected COVID-19 patients
Journal of Radiation Protection and Research ; 46(3):120-126, 2021.
Article in English | Scopus | ID: covidwho-1502825
ABSTRACT

Background:

This paper aims to evaluate the clinical utility and radiation dosimetry, for the mobile X-ray imaging of patients with known or suspected infectious diseases, through the window of an isolation room. The suitability of this technique for imaging coronavirus disease 2019 (COVID-19) patients is of particular focus here, although it is expected to have equal relevance to many infectious respiratory disease outbreaks. Materials and

Methods:

Two exposure levels were examined, a “typical” mobile exposure of 100 kVp/1.6 mAs and a “high” exposure of 120 kVp/5 mAs. Exposures of an anthropomorphic phantom were made, with and without a glass window present in the beam. The resultant phantom images were provided to experienced radiographers for image quality evaluation, using a Likert scale to rate the anatomical structure visibility. Results and

Discussion:

The incident air kerma doubled using the high exposure technique, from 29.47 μGy to 67.82 μGy and scattered radiation inside and outside the room increased. Despite an increase in beam energy, high exposure technique images received higher image quality scores than images acquired using lower exposure settings.

Conclusion:

Increased scattered radiation was very low and can be further mitigated by ensuring surrounding staff are appropriately distanced from both the patient and X-ray tube. Although an increase in incident air kerma was observed, practical advantages in infection control and personal protective equipment conservation were identified. Sites are encouraged to consider the use of this technique where appropriate, following the completion of standard justification practices. Copyright © 2021 The Korean Association for Radiation Protection.