End-Tidal Carbon Dioxide Pressure Measurement after Prolonged Inspiratory Time Gives a Good Estimation of the Arterial Carbon Dioxide Pressure in Mechanically Ventilated Patients.
Salomé, Arthur; Stoclin, Annabelle; Motamed, Cyrus; Sitbon, Philippe; Bourgain, Jean-Louis.
  • Salomé A; Assistance Publique des Hôpitaux de Paris, 94270 Le Kremlin Bicêtre, France.
  • Stoclin A; Faculty of Medicine, University of Paris-Saclay, 94270 Le Kremlin Bicêtre, France.
  • Motamed C; Intensive Care Unit, Gustave Roussy Institute, 94800 Villejuif, France.
  • Sitbon P; Intensive Care Unit, Gustave Roussy Institute, 94800 Villejuif, France.
  • Bourgain JL; Anesthesia Department, Gustave Roussy Institute, 94800 Villejuif, France.
Diagnostics (Basel) ; 11(12)2021 Nov 27.
Article in English | MEDLINE | ID: covidwho-1542452
ABSTRACT

BACKGROUND:

End-tidal carbon dioxide pressure (PetCO2) is unreliable for monitoring PaCO2 in several conditions because of the unpredictable value of the PaCO2-PetCO2 gradient. We hypothesised that increasing both the end-inspiratory pause and the expiratory time would reduce this gradient in patients ventilated for COVID-19 with Acute Respiratory Distress Syndrome and in patients anaesthetised for surgery.

METHODS:

On the occasion of an arterial blood gas sample, an extension in inspiratory pause was carried out either by recruitment manoeuvre or by extending the end-inspiratory pause to 10 s. The end-expired PCO2 was measured (expiratory time 4 s) after this manoeuvre (PACO2) in comparison with the PetCO2 measured by the monitor. We analysed 67 Δ(a-et)CO2, Δ(a-A)CO2 pairs for 7 patients in the COVID group and for 27 patients in the anaesthesia group. Results are expressed as mean ± standard deviation.

RESULTS:

Prolongation of the inspiratory pause significantly reduced PaCO2-PetCO2 gradients from 11 ± 5.7 and 5.7 ± 3.4 mm Hg (p < 0.001) to PaCO2-PACO2 gradients of -1.2 ± 3.3 (p = 0.043) and -1.9 ± 3.3 mm Hg (p < 0.003) in the COVID and anaesthesia groups, respectively. In the COVID group, PACO2 showed the lowest dispersion (-7 to +6 mm Hg) and better correlation with PaCO2 (R2 = 0.92). The PACO2 had a sensitivity of 0.81 and a specificity of 0.93 for identifying hypercapnic patients (PaCO2 > 50 mm Hg).

CONCLUSIONS:

Measuring end-tidal PCO2 after prolonged inspiratory time reduced the PaCO2-PetCO2 gradient to the point of obtaining values close to PaCO2. This measure identified hypercapnic patients in both intensive care and during anaesthesia.