Predicting flow diverter sizing using the AneuGuideTM software: a validation study.
Tong, Xin; Shan, Yejie; Leng, Xiaochang; Chen, Jigang; Fiehler, Jens; Siddiqui, Adnan H; Hu, Xuebin; Liu, Aihua; Xiang, Jianping.
  • Tong X; Neurointervention Center, Beijing Neurosurgical Institute and Beijing Tiantan Hospital, Beijing, China.
  • Shan Y; ArteryFlow Technology Co., Ltd, Hangzhou, China.
  • Leng X; ArteryFlow Technology Co., Ltd, Hangzhou, China.
  • Chen J; Neurointervention Center, Beijing Neurosurgical Institute and Beijing Tiantan Hospital, Beijing, China.
  • Fiehler J; Department of Neuroradiology, University Medical Center Hamburg Eppendorf, Hamburg, Germany.
  • Siddiqui AH; Neurosurgery and Radiology and Canon Stroke and Vascular Research Center, University at Buffalo Jacobs School of Medicine and Biomedical Sciences, Buffalo, New York, USA.
  • Hu X; Department of Neurosurgery, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China.
  • Liu A; Neurointervention Center, Beijing Neurosurgical Institute and Beijing Tiantan Hospital, Beijing, China liuaihuadoctor@163.com jianping.xiang@arteryflow.com.
  • Xiang J; ArteryFlow Technology Co., Ltd, Hangzhou, China liuaihuadoctor@163.com jianping.xiang@arteryflow.com.
J Neurointerv Surg ; 2022 Jan 17.
Article in English | MEDLINE | ID: covidwho-2214004
ABSTRACT

BACKGROUND:

Stent sizing remains a challenging task for flow diverter implantation because of stent foreshortening. In this study, we aimed to quantify the change in length after implantation and assess the error in length prediction using AneuGuideTM software.

METHODS:

In a retrospective cohort of 101 patients with 102 aneurysms undergoing treatment with a pipeline embolization device (PED; Covidien, Irvine, California, USA), we used AneuGuideTM software to obtain measured lengths (ML) and calculated lengths (CL) after stent implantation. Stent elongation was defined as the ratio of ML-LL to the labeled length (LL). Simulation error was defined as the ratio of the absolute value of CL-ML to ML. The correlation and consistency between ML and LL and between ML and CL were analyzed using Pearson's correlation test and the Bland-Altman plot. Statistical significance was set at p<0.05.

RESULTS:

The mean elongation of ML was 32.6% (range 26.3-109.2%). Moderate consistency was observed between LL and ML (ρ=0.74, p<0.001). With the AneuGuideTM software, the mean simulation error was 6.6% (range 0.32-21.2%). Pearson's correlation test and the Bland-Altman plot showed a high correlation and consistency between ML and CL (ρ=0.96, p<0.001).

CONCLUSION:

Labeled length provides only a low reference value for predicting the actual length of the flow diverter after implantation. The high consistency between ML and CL obtained from AneuGuideTM software shows its great potential for the optimization of the flow diverter sizing process.