Your browser doesn't support javascript.
Real-world Evaluation of a Sample Pooling Strategy for Large-Scale Rapid COVID-19 Testing.
Lee, Hao-Ru; Wang, Fang-Yu; Li, Jia-Ling; Chou, Teh-Ying; Ho, Hsiang-Ling.
  • Lee HR; Department of Pathology and Laboratory Medicine, Taipei Veterans General Hospital, 201, Section 2, Shipai Road, Taipei 11217, Taiwan.
  • Wang FY; Department of Pathology and Laboratory Medicine, Taipei Veterans General Hospital, 201, Section 2, Shipai Road, Taipei 11217, Taiwan.
  • Li JL; Department of Pathology and Laboratory Medicine, Taipei Veterans General Hospital, 201, Section 2, Shipai Road, Taipei 11217, Taiwan.
  • Chou TY; Department of Pathology and Laboratory Medicine, Taipei Veterans General Hospital, 201, Section 2, Shipai Road, Taipei 11217, Taiwan.; Department of Biotechnology and Laboratory Science in Medicine, National Yang-Ming University, Taipei, Taiwan.
  • Ho HL; Department of Pathology and Laboratory Medicine, Taipei Veterans General Hospital, 201, Section 2, Shipai Road, Taipei 11217, Taiwan.; Department of Biotechnology and Laboratory Science in Medicine, National Yang-Ming University, Taipei, Taiwan. Electronic address: hlho5@vghtpe.gov.tw.
J Clin Virol ; 149: 105133, 2022 04.
Article in English | MEDLINE | ID: covidwho-1739902
ABSTRACT

BACKGROUND:

The worldwide outbreak of COVID-19 has become a public health crisis of unprecedented proportions. The fast spread of emerging variants increases the needs of rapid diagnostic and screening testing. Sample pooling efficiently expands the testing capacity under limited resources.

OBJECTIVES:

We evaluated the performance of sample pooling on the Point-of-Care (POC) Liat® and cobas® 6800 systems and provided real-world experiences for implementing these systems in large-scale screenings.

METHODS:

Positive nasopharyngeal (NP) specimens with Ct values < 25, 25∼30 or > 30 were tested individually and in pools to optimize the POC Liat® and cobas® 6800 systems, which were then implemented in community screenings.

RESULTS:

The 5-sample pooling strategy did not affect the positive detection rates on Liat® or cobas® 6800 in samples with Ct values <25 or 25∼30. However, in samples with low viral loads (Ct values >30), five-sample pooling has a higher positive detection rate on POC Liat® (20/20; 100%), compared to cobas® 6800 (9/20; 45%). Five-sample pooled on POC Liat® and two-sample pooled on cobas® 6800 appear to be appropriate for SARS-CoV-2 detection. By implementing the pooling strategies in two large-scale community screenings, 7,606 NP specimens was tested within 36 h; the average turn-around time was 4.8 h for cobas® 6800 and 1.3 h for POC Liat®. Eight positive specimens (0.11%; 8/7,606) were identified, with Ct values ranging from 18.85 to 37.68.

CONCLUSION:

The performance of sample pooling on POC Liat® was demonstrated to be an effective, accurate, and economical approach for large-scale community screenings for COVID-19.
Subject(s)
Keywords

Full text: Available Collection: International databases Database: MEDLINE Main subject: COVID-19 Testing / COVID-19 Type of study: Diagnostic study / Experimental Studies / Qualitative research Topics: Variants Limits: Humans Language: English Journal: J Clin Virol Journal subject: Virology Year: 2022 Document Type: Article Affiliation country: J.jcv.2022.105133

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Collection: International databases Database: MEDLINE Main subject: COVID-19 Testing / COVID-19 Type of study: Diagnostic study / Experimental Studies / Qualitative research Topics: Variants Limits: Humans Language: English Journal: J Clin Virol Journal subject: Virology Year: 2022 Document Type: Article Affiliation country: J.jcv.2022.105133